IVT-LVS630B(200W) Ligasure Electrosurgical Unit
IVT-LVS630B(200W) Ligasure Electrosurgical Unit with 7‑inch touch panel, permanent vessel sealing up to 7 mm, REM safety protection, real‑time tissue‑impedance feedback & 99 user preset memory slots.
PRODUCT DETAILS
Technical Features:
Intelligent Device---7 INCH LCD TOUCH SCREEN
--The intuitive design and user-friendly set-up of Electrosurgical Unit in the modern operating room,
--Monopolar and bipolar functions, It can satisfy all the surgeries demands, with safety, flexibility, reliability and convenience.
--With a wide TFT LCD touch screen (7 inch), clean and sharper image quality, which provides the user with easy access to all functions. The settings or modes of operation are changed by touching icons on the screen to ensure maximum ease of operation.
REM (Return Electrode Monitoring)
The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimize the danger of burning incidents.

Automatic Self-Test
Once switched on, Electrosurgical Systems perform a comprehensive internal test
Real-time Monitoring and Instance Response System for Tissue Density
This proprietary technology delivers optimal clinical effects through continuous synchronization of current and voltage. It samples the current and voltage 450,000 times per second which enables it to respond to tissue impedance changes in less than 10 milliseconds, which ensuring the machine achieves optimum energy output levels quicker and more accurately – ensuring that only the precise voltage needed is safely delivered to each tissue
type

Vessel Sealing
The machine enables permanent sealing up blood vessels with a diameter up to 7mm with this technology under bipolar coagulation..
Technical Specification
Main Frequency: 512 KHz Power Supply: 100-240V/50-60Hz
Weight and Dimension:
Weight:8.00KG Dimension: 47*34*17
High Frequency Leakage:
Monopolar: < 150mA; Bipolar <60mA
Low Frequency Leakage:
Normal Polarity, Intact Chassis Ground: <10μA
Normal Polarity, Ground open< 50μA
Reverse Polarity, Ground Open< 50μA
Sink Current, 140V Applied, All inputs < 50μA
Technical Specification
| Working mode | Max.output power | Rated Load | |
| Monopolar Cut | Pure Cut | 200 W | 500 Ω |
| Blend CutⅠ | 180 W | 500 Ω | |
| Blend Cut Ⅱ | 150 W | 500 Ω | |
| Monopolar Coagulation | Soft Coagulation | 150 W | 500 Ω |
| Fulgurate Coagulation | 120 W | 500 Ω | |
| Spray Coagulation | 100 W | 500 Ω | |
| Bipolar Coagulation | Standard Coagulation | 150 W | 100 Ω |
| Precise Coagulation | 60 W | 100 Ω | |
| Ligasure (Vessel Sealing) | 200 W | 25 Ω | |